DIRECTION 5854713-100, REVISION 2 VENUE FIT™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-41
PRELIMINARY
8-3-3 Software Installation Procedure
NOTE: Images in this procedure are for reference only. There may be differences or variations, in accordance
with different software versions.
1.) Plug the Software Installation Media into the USB port located on the Interface Panel of the Venue
Fit™ .
2.) Turn on the system.
The opening screen is displayed.
3.) For new installation (after SSD replacement), tap Format SSD and Install.
If patient information is stored on the SSD, perform one of the following:
Select Update SW to install software without affecting user data.
Figure 8-63 Venue Fit™ Software - Installation Options
The Warning message appears:
Figure 8-64 Venue Fit™ SSD Format- Warning Message
4.) Click Yes in the Warning message box.

Table of Contents