VENUE FIT™ SERVICE MANUAL
5-18 Chapter 5 - Components and Function (Theory)
PRELIMINARY
Section 5-10
Cooling System
5-10-1 General Information
The Venue Fit™ system has 2 fans for system cooling:
To allow efficient cooling, the Scanner cooling consists of the following elements:
Heat Pipe to transfer the heat (installed on the C-BEB).
Outlet Blower at the edge of the Heat Pipe.
Fan to push cold air into scanner (installed on the C-FEB Heat Sink).
All fans are controlled and have two variable speeds.
The cooling requirement for the Venue Fit™ ultrasound scanner with on board peripherals, is up to 600
BTU/h. This figure does not include cooling needed for lights, people, or other equipment in the room.
NOTE: Each person in the room places an additional 300 BTU/h demand on the cooling system.
Figure 5-3 Air Flow