DIRECTION 5854713-100, REVISION 2 VENUE FIT™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-11
PRELIMINARY
8-2-6 M-PSB Replacement Procedure
8-2-6-1 Tools
Phillips screwdriver of appropriate size
8-2-6-2 Time Required
5 min
8-2-6-3 Preparations
1.) Shut down the Venue Fit™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-6-4 M-PSB Removal Procedure
1) Remove Rear cover.
2) Remove Speakers.
3) Release 2 captive screws from the top side of the M-PSB.
4.) Gently Pull upwards the M-PSB module.
Figure 8-39 M-PSB Removal
Rear Cover Removal Procedure
Speakers Removal Procedure