DIRECTION 5854713-100, REVISION 2 VENUE FIT™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-49
PRELIMINARY
The selected data is copied. The backup process is indicated on the progress bar that appears.
6.) After the backup is complete, proceed to Software Update Procedure.
8-3-4-3 Software Update Procedure
1.) Turn OFF the system.
2.) Plug the Software Update Media into the USB port located on the Interface Panel of the Venue
Fit™.
3.) Turn on the system.
The opening screen is displayed.
4.) For Software Update: Select Update Venue Fit™ SW to upgrade software without affecting user
data.
Figure 8-82 Venue Fit™ Software - Update Options
Figure 8-81 Copying Patients Dialog