DIRECTION 5854713-100, REVISION 2 VENUE FIT™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-5
PRELIMINARY
8-2-3 Kickstand Installation Procedure
8-2-3-1 Tools
Phillips screwdriver of appropriate size.
8-2-3-2 Time Required
5 min
8-2-3-3 Preparations
1.) Shut down the Venue Fit™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-3-4 Kickstand Installation Procedure
1) Remove the hinges rubber caps by pushing them to the sides and dispose them.
Figure 8-33 Rubber caps removal
2) Align the kickstand below the hinges and secure with 2 screws on each side.
Figure 8-34 Kickstand installation