GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 3 - System Setup 3-35
3-7-2-2 Wireless USB Network Adaptor
• No SafeLock Cart - connection directly to the Vivid q N system
The installation procedure requires connecting the DVD-CD drive. The procedure includes installing
the Compact Wireless-G USB Adaptor software and then the configuration procedure, which is
performed via the Windows Network Connections support screens.
For procedure instructions, see "Wireless USB Network Configuration" on page 3-90.
For network security reasons, when configuring the wireless USB network adaptor on the
Vivid q N system, do not use the USB ports on the SafeLock cart.
For network security reasons, both USB ports on the Vivid q N system must be configured
separately when installing the wireless USB network adaptor.
Do not use any other type of wireless adaptor other than the LinkSys WSUB54GC WLAN card.
The ultrasound system is an extremely sensitive and complex medical system. Any
unauthorized peripherals may cause system failure or damage!
Figure 3-28 Wireless USB Network Adaptor Connected to Rear of Vivid q N Scanner
R-Adaptor
Wireless USB Network Adaptor