Chapter 8 - Replacement Procedures 8-49
Figure 8-57 below.
5.) Carefully lift the DC power supply upwards and remove it from the Bottom Assembly, as shown in
Figure 8-58 below.
6.) Proceed to perform the following functionality tests:
- "SafeLock Cart - Grounding Continuity" on page 10-22 - only if the Vivid q N system is
currently mounted on the SafeLock cart
- "SafeLock Cart - Chassis Current Leakage Test" on page 10-24
- "Calibration Tests" on page 7-25
- "Full System Test" on page 7-10
Figure 8-57 Loosening the Screws on the DC Power Supply
Figure 8-58 Removing the DC Power Supply