IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8-78 Section 8-4 - SafeLock Cart Components Replacement
6.) Continue routing the Peripheral AC Outlet cable towards the right hand side of the Rear Upper
Cover and downwards - Figure 8-92.
Figure 8-92 Routing of Cable Downwards inside Rear Upper Cover
7.) Fasten the Peripheral AC Outlet cable to the Main Harness cable as shown in Figure 8-93.
Figure 8-93 Peripheral AC Outlet Cable Fastened to Main Harness Cable
8.) Plug the Peripheral AC Outlet cable into the power supply and connect the Ground lead to the body
of the cart - Figure 8-94.
Note use of cable fasteners