OMGA5KN4.4 GA5000 gas analyser
Page 24 of 113 Copyright Geotech
please take this into account when estimating battery life.
Note: Connect the charger to the mains attaching the appropriate adaptor.
Power supply front and back drawing:
Charger: Input voltage: 100-240V AC +/- 10%
Input frequency: 50-60Hz +/- 10%
Input current: 0.4A@100VAC .. 0.2A@240VAC
Output voltage: 10.1VDC max
Output current: 1.5A max
Note: This charger has been internally restricted to 1.5A
6.10 Cleaning instructions
Do NOT use any cleaning agents to clean the analyser or battery charger as they may
have an adverse effect on the safe use of these devices.
6.11 Memory
The analyser's memory is stored in a readings and configuration database. The analyser
will prompt when its memory is full, and you will not be able to store any further readings.
Please download your readings via GAM or the Basic Download Software and then clear the
memory.
Note: The analyser should never be stored for prolonged periods with valuable data in
its memory. It is advisable to download all readings to GAM at the end of each
day’s monitoring. To clear the memory, please refer to the Gas Analyser
Manager (GAM) operating manual.