EasyManuals Logo

Geotech GA5000 User Manual

Geotech GA5000
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
GA5000 gas analyser OMGA5KN4.4
Copyright Geotech Page 43 of 113
Adjust flow fail
3) Manual adjustment of the flow fail is available via this option and can be carried out
with use of ‘Key 4 – Scroll left’ () less sensitive and ‘Key 6 – Scroll right’ (>) more
sensitive.
4) Select the soft-key Save to store the setting or select soft-key Exit’ to exit the screen
without saving the change.
5) The operator will return to the ‘Operation settings’ menu.
Note: The default setting displays the bar in the centre. BEFORE altering this
setting, please contact Technical Support at Geotech on +44(0)1926
338111 or email technical@geotech.co.uk
7.4.6 Technician login
This option enables the operator to select or change a pre-defined
technician login and all subsequent readings will be tagged with this
Technician Login ID. The technician ID must already have been created
using the Gas Analyser Manager (GAM) software and uploaded to the
instrument.
1) Select the ‘Menu’ button on the front of the analyser to display the ‘Device Settings’
menu. Press the soft key to display ‘User Settings’.
2) Select ‘Key 6 – Technician login’ and the following screen is displayed:
Technician login

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Geotech GA5000 and is the answer not in the manual?

Geotech GA5000 Specifications

General IconGeneral
BrandGeotech
ModelGA5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals