| CARDIOHELP System | 3 Inspection | 47 |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
3 Carry out the touchscreen calibration (⇨ "Calibrate the
Touchscreen", page 78).
Calibrating the touchscreen
u Rotary knob button function - passed/failed.
u Touchscreen calibration completed.
Emergency Mode Button Function
1 Press the [Emergency mode] button for 4 s.
The emergency mode control panel opens.
2 Turn the rotary knob
3 Verify that the set rpm changes.
4 Press the [Emergency mode] button for 10 s to leave the function.
5 Restart the device by pressing the on/off button.
u Emergency mode button function - passed/failed.
Unlock Function
1 Press the [Unlock] button.
[Unlock] LED lights up.
[Locked] symbol
The [Locked] symbol appears in the status bar.
2 Touch the touchscreen.
"System locked"
The message [System locked] appears.
3 Unlock the system according to the instructions on the touchscreen.
[Unlock] LED goes out.
The [Not locked] symbol appears in the status bar.
[Not locked] symbol
u Unlock function - passed/failed.

Table of Contents

Other manuals for Getinge MAQUET CARDIOHELP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge MAQUET CARDIOHELP and is the answer not in the manual?

Getinge MAQUET CARDIOHELP Specifications

General IconGeneral
BrandGetinge
ModelMAQUET CARDIOHELP
CategoryMedical Equipment
LanguageEnglish

Related product manuals