| 46 | 3 Inspection | CARDIOHELP System |
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
▶ The battery status window is opened.
▶ Both battieries are detected and the capacity is displayed.
u Battery detection / battery capacity - passed/failed.
3 Connect the CARDIOHELP to the external power supply with the
AC power supply cord.
Battery status window with ext. power supply
▶ Green AC/DC power supply LED must light up.
▶ Green battery LED must light up.
NOTE
Remaining battery capacity must be <95%!
CAUTION!
Carry out the following tests without external power supply if the green battery LED is not lit and the
remaining battery capacity is thus over 95%.
Repeat the “Battery LED and AC/DC power supply LED function” test after approx. 30 minutes.
▶ Arrow and mains plug symbol are displayed in the “Battery” status window.
u AC power supply function - passed/failed.
u Battery LED function and AC/DC power supply LED passed/failed.
4 Connect the CARDIOHELP to the external power supply with the DC power cord.
▶ Green AC/DC power supply LED must light up.
▶ Green battery LED must light up.
NOTE
Remaining battery capacity must be <95%!
▶ Arrow and mains plug symbol are displayed in the “Battery” status window.
u DC power supply function - passed/failed.
Rotary Knob Button Function
1 Press the [Safety button] and hold it down.
2 Press the [Rotary knob] button for min. 10 s.
▶ The touchscreen calibration window opens.
Control panel
NOTE
Speed must be set to 0 rpm.