86
4. electrIcal connectIon
GH2 is equipped with batteries that require regular
recharging.
Class I equipment
Fixed rail systems are class I equipment and must be
installed by a qualified installer or by Guldmann Service
Team.
Equipment is disconnected from Supply Mains by breaking
the main breaker switch.
Class II equipment
Mobile equipment is class II equipment (marked with the
double-encased symbol) and can be connected to the
mains direct by the user.
Charger Mascot type 2240 (DC systems) must be used.
Equipment is disconnected from Supply Mains by detach
the mains plug from the wall outlet.
Safety concerning static electricity (ESD)
Service technicians and installers must use an ESD-safety
package consisting of a mat, a ground wire, and a
bracelet. The technician/installer connects the mat to a
grounding point, for instance a radiator or a water pipe.
He must then put on the bracelet and connect it to the mat.
If it is not possible to find a grounding point, the mat and
the bracelet must be used as a minimum.
Only then is it allowed to work with the PC Board or
components where it is possible to come into contact with
the PC Board.