REMOTE CONTROL
197
CALCulate:STORe:COUNt
Set the number of measurement counts that are recorded with the Store
measurement function.
Parameter: <NR1> (2 ~ 9999) | MIN | MAX
Example: CALC:STOR:COUN 1000
Sets the number of counts to be recorded as 1000.
CALCulate:STORe:COUNt?
Returns the number of counts that are recorded with the Store
measurement function.
Parameter: [None] | MIN | MAX
CALCulate:AVERage:COUNt
Sets the total number of statistic counts.
Parameter: <NR1> (0, 2~100000) 0=continuous count, 2~100000=count
Example: CALC:AVER:COUN 0
Sets the count to continuous.
CALCulate:AVERage:COUNt?
Returns the total number of recorded counts. The setting commands for
this query are: CALCulate:STORe:COUNt, ROUTe:COUNt and
CALCulate:AVERage:COUNt.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats
Example: CALC:AVER:COUN? 0
>+0010
Returns the total number of counts set for the Store function (10 counts).
CALCulate:AVERage:MINimum?
Returns the minimum recorded value.
Parameter: None | <NR1>(0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:MAXimum?
Returns the maximum recorded value.
Parameter: None | <NR1>(0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:AVERage?
Returns the average recorded value.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats
CALCulate:AVERage:PTPeak?
Returns the recorded peak to peak value (max value min value).
Parameter: None | <NR1> (0|1|2) 0=Store, 1=Scan, 2=Stats
Return Parameter: <NRf>
CALCulate:AVERage:SDEViation?
Returns the recorded Standard Deviation.
Parameter: None | <NR1> (0~2) 0=Store, 1=Scan, 2=Stats