GDM-8351 User Manual
146
DC Voltage
Range
[1]
Resolution
Full Scale
Accuracy
100.000mV
0.001mV
120.000
0.012% + 8
1.00000 V
0.00001V
1.20000
0.012% + 5
10.0000 V
0.0001V
12.0000
0.012% + 5
100.000 V
0.001V
120.000
0.012% + 5
1000.00 V
[2]
0.01 V
1020.00
0.012% + 5
[1] When the input value exceeds the full scale of the selected range, the display
will show -OL- (over load) on the display.
[2] The specifications are guaranteed to an input voltage of 1000V. A beeping
alarm will go off when the input voltage is higher than 1000V.
DC Current
Range
[1]
Resolution
full scale
Accuracy
10.0000mA
0.0001mA
12.0000
0.05% + 15
100.000mA
0.001mA
120.000
0.05% + 5
1.00000A
0.00001 A
1.20000
0.2% + 5
10.0000A
[2]
0.0001 A
12.0000
0.2% + 5
[1] When the input value exceeds the full scale of the selected range, the display
will show -OL- (over load) on the display.
[2]The specifications are guaranteed to an input of 10A. A beeping alarm will go
off when the input value is higher than 10A.
Diode
Test voltage
Resolution
Maximum reading
Accuracy
6V
0.0001 V
5.9999V
0.012% + 5
* The diode test voltage is 6V, 1mA.
Continuity
Range
Resolution
Maximum reading
Accuracy
1000.00
0.01Ω
1200.00
0.1% + 8
* Without REL function, add 0.2 Ω additional error.