APPENDIX
147
Resistance
[1]
[2]
[1] Specifications are for 4-wire resistance measurement, or 2-wire resistance
measurement using the REL function. Without the REL function, add 0.2 Ω
additional error when using 2-wire resistance measurement.
[2] When measuring resistances greater than 500kΩ, please use shielded test
leads to eliminate the noise interference that may be induced by standard test
leads.
[1] Specifications are for sine wave inputs that are greater than 5% range.
[2] Rate in Fast, Input ACV Frequency > 200Hz.
[3] The specifications are guaranteed to an input of 750V. A beeping alarm will
go off when the input value is higher than 750V.