APPENDIX
149
Thermocouple Specifications
Type
Measurement
Range
Resolution
Accuracy
(-200 ~ 0°C)
Accuracy
(0 ~ 300°C)
J, K, T
-200 ~ +300°C
0.01°C
0.4°C
0.2°C
*Specifications do not include probe accuracy.
Capacitance
Range
Resolution
Full Scale
Test Current
Accuracy
10.00nF
[1]
0.01nF
12.00
10μA
2.0%+10
100.0nF
0.1nF
120.0
10μA
2.0%+4
1.000μF
0.001μF
1.200
100μA
2.0%+4
10.00μF
0.01μF
12.00
1mA
2.0%+4
100.0μF
0.1μF
120.0
1mA
2.0%+4
*Specifications are for film Capacitance inputs that are greater than 10% range.
[1]10nF capacitance measurements may be affected by the stray capacitance on
the test cables. Before testing, use the REL function to compensate for the stray
capacitance from the test cables.