WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 101
Anatel Operator Manual
Cancel icon returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
2) Close the bottle bay door.
3) Touch the forward icon. The excursion mode setup dialog box appears. See
Figure 6-30.
Figure 6-30 Excursion mode setup dialog box
TOC limit text box allows you to enter a limit an online TOC analysis must reach
before an excursion sample is captured. You may specify a value from 1 to 2000
ppb. The default setting is 500 ppb.
Run now icon immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See Figure 6-27. Once the
bottle has been filled, the excursion captured dialog box appears. See Figure 6-28,
page 99.
Done icon returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
Cancel icon takes you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.