3.1 Simple Test Operations
This section describes the Simple Test operation. The Simple Test
generates an (adjustable rate) high voltage ramp until a breakdown
happens. The user may repeat the procedure for one insulator-oil filling and
that will be considered as a single test. The possible settings are the rate of
rise for the voltage and the stirrer control. At the end of the test the system
will prompt the user for a printout and the test results will be stored in the
History structure (see paragraph xxx).
The flowchart in figure 3-1 depicts the operations of the Simple Test.
Figure 3-
1 – Simple
Test Flowchart
3.2
Simple
Test
Initializ
ation
After the
user selects
the execution
of a Simple
Test the
system
displays a
warning
message for
the electrodes
spacing and
the system is
halted, Figure
3-2.
SIMPLE TEST
Get the rate of rise
Compute the desired
voltage ramp
BreakDown
Interrupt
Output
Voltage
Compute Accurate
High Voltage
Another test with
the same sample?
Verify Electrodes/Gap
Spacing
Wait for Enter
Key
Print Report
Voltage
Measurement
Voltage
Controller
Breakdown
Detector
Concurrent Processes
LCD Update
Voltage
Wait for
Breakdown
Interrupt