EasyManua.ls Logo

High Voltage DTS-60A - Page 53

High Voltage DTS-60A
115 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTS-60A OPERATORS MANUAL
43
large digits indicate the current high voltage output (in kV) and the left icon
indicates the Stirrer State (on/off). Finally, the last line is used for user
interface messages, such as the time/date/temperature message.
If the system detects a breakdown, it will display the event, Figure 4-9,
and the execution will proceed according to the Test Program commands.
Figure 4-9 – Breakdown
Finally, if the Test terminates a respective system message will be
displayed, Figure 4-10. The system prompts the user for a printout. Finally,
the system will return to the initial state, after pressing the ‘ENTER’ key,
Figure 4-7.
1.0kV/s > BREAKDOWN
DURATION:
30.7
00:00:31
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
TEST:IEC-156/95
12:10 01/08/05 65 °F
STAGE
1
HV on
Current Stage/ High Voltage ON Indicator
AFTER BD WAIT TIMER...
DURATION:
42.2
00:01:00
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
TEST:IEC-156/95
12:10 01/08/05 65 °F
Current Stage Indicator
STAGE
1
Blinking Breakdown Indicator