DTS-60A OPERATORS MANUAL
43
large digits indicate the current high voltage output (in kV) and the left icon
indicates the Stirrer State (on/off). Finally, the last line is used for user
interface messages, such as the time/date/temperature message.
If the system detects a breakdown, it will display the event, Figure 4-9,
and the execution will proceed according to the Test Program commands.
Figure 4-9 – Breakdown
Finally, if the Test terminates a respective system message will be
displayed, Figure 4-10. The system prompts the user for a printout. Finally,
the system will return to the initial state, after pressing the ‘ENTER’ key,
Figure 4-7.
1.0kV/s > BREAKDOWN
DURATION:
30.7
00:00:31
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
TEST:IEC-156/95
12:10 01/08/05 65 °F
STAGE
1
HV on
Current Stage/ High Voltage ON Indicator
AFTER BD WAIT TIMER...
DURATION:
42.2
00:01:00
Test Program Name
Count up Timer for current stage duration
Rate of rise from 0kV to breakdown
System Messages
kV
TEST:IEC-156/95
12:10 01/08/05 65 °F
Current Stage Indicator
STAGE
1
Blinking Breakdown Indicator
STIR