DTS-60A OPERATORS MANUAL
62
The items in figure 4-12 can get the following values:
Command = new
1. Program Name
Up to 12 characters. Do not use space. Use ‘_’ instead.
2. Electrode Gap Value
Floating-point value that indicates the Electrode Spacing (eg.
Electrode Spacing = 2.5 +/- 0.01 mm)
3. Electrode Gap Error
Floating-point value that indicates the Electrode Spacing possible
error (eg. Electrode Spacing = 2.5 +/- 0.01 mm)
4. Electrode Gap Units
Units defined for the above values.
Options: mm or in (eg. Electrode Spacing = 2.5 +/- 0.01 mm)
5. Electrode Type
Options: Type_I, Type_II, Type_I_II
6. Startup WAIT Timer -SWT
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)
7. Startup STIRRER Timer - SST
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)
8. Number of Stages
Integer value (0…29)
Command = stage
1. Stage ID
Indicates the stage that the following information refers.
Integer value (0…29)
2. Voltage Value
Floating-point value that indicates the desired Voltage
3. Voltage Type
Options: kV/s, kV
4. Voltage Duration
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)
5. Stirrer Timer- ST
The time that the stirrer will be enabled (in respect with the
above timer).
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)
6. After Breakdown WAIT Timer- ABDWT
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)
7. After Breakdown STIRRER Timer - ABDST
hh:mm:ss (hh = hours, mm = minutes, ss = seconds)