167
_____________________________________________________________________________________________
7.3 Time Taken for Testing
______________________________________________________________________________________________
<B>
The time taken for test varies according to the display state
(display ON/OFF):
Display setting Test time
LCD display and backlight :OFF -2.3 ms
Voltage and current monitors:OFF -1.4 ms
LCD display and backlight :ON
Voltage and current monitors:ON
0ms
Allowance 1ms
<C>
The time taken for calculation varies according as to whether or not both
open circuit compensation and also short circuit compensation are performed:
Open/short circuit compensation Calculation time
Open and/or short circuit
compensation performed
1.5 ms
Not performed 0ms
Allowance 1ms
<D>
The time taken for calculation varies according as to whether or not the
comparator is operating:
Comparator operating Calculation time
Normal testing 0ms
When the comparator is operating 0.7 ms
Allowance 1ms
<E>
The time taken for calculation varies according as to whether or not the
comparator is operating:
Scaling execution Calculation time
Normal testing 0ms
When the scaling is executed 0.8 ms for a parameter
Allowance 1ms