4.6 Setting the Ranging
______________________________________________________________________________________________
NOTE
4.6.1 Control Screen Sequence
Initial screen
Ranging setting screen
Set to HOLD
Range setting is
performed manually
Initial screen
Menu screen
Set to AUTO
Range setting is
performed automatically
4.6 Setting the Ranging
Using AUTO, the 3532-50 unit can automatically select the most appropriate
test range. This is useful when testing a sample whose impedance varies
greatly with frequency, or whose nature is unknown. The test range can also
be fixed using HOLD, which provides the quickest testing.
AUTO :
The most suitable test range is set automatically.
HOLD :
The test range is fixed, and may only be altered manually.
The ranges are all defined in terms of impedance. Therefore, for a parameter
other than impedance, the value is obtained by calculating from the
measured values of |Z|and
θ
.
Press the key.
Press the key.
Choose the manner in which the range will be set
(AUTO or HOLD).
When the ranging setting has been established as
above, press the key to return to the Initial
screen.