4.5 Setting the Voltage/Current Limit
______________________________________________________________________________________________
4.5.2 Details of the Setting Process
Initial screen
Menu screen
Monitor display
Voltage Limit Screen
Current Limit Screen
Limit ON/OFF key
When the test signal level type
is voltage (V or CV)
When the test signal level type
is current (CC)
NOTE
Test signal control level Limit set Setting range
V, CV Current limit 0.01 mA to 99.99 mA
CC Voltage limit 0.01 V to 5 V
NOTE
1. Press the key on the Initial screen to display the Menu screen.
2. Press the key on the Menu screen to display the one of the Limit
setting screens.
The present situation can be checked using the monitor display.
The monitor display is different for V, CV, and CC.
The setting for voltage or current limit changes automatically to current or
voltage limit, according to the present Test signal level setting. Therefore,
first set the test signal level, and thereafter set the voltage or current limit.
For details of the Test signal level setting, refer to Section 4.4.
3. Input the limit value using the digit keys. The ranges within which the limit
values can be set are as follows.
The accuracy of the current limit is 10% 10μA for 42.0 Hz to 1.000
MHz, and 20% 10μA for 1.001 MHz to 5.000 MHz.
The accuracy of the voltage limit is 10% 10 mV for 42.0 Hz to 1.000
MHz, and 20% 10 mV for 1.001 MHz to 5.000 MHz.
When any digit key would take the limit value outside these ranges, it
changes to non-reversed video (black characters upon a white ground).
The limit value is not made effective just by setting it, if the limit function
is turned off.