173
_____________________________________________________________________________________________
7.5 Test Accuracy
______________________________________________________________________________________________
Z
θ
RANGE
Basic Accuracy Coefficient Table
(0.501V to 1.000V)
A=0.08 B=0.01
A=0.05 B=0.005
1.001kHz - 10kHz
1kΩ
0.01 x |10 x 500 - 1 x 10
3
|
Z accuracy =
(0.08 +
______________________
)= 0.12%
1x 10
3
0.005 x |10 x 500 - 1 x 10
3
|
θ accuracy =
(0.05 +
______________________
)= 0.07
1x 10
3
Z
θ
RANGE
Basic Accuracy Coefficient Table
(0.501V to 1.000V)
A=0.08 B=0.01
A=0.05 B=0.005
100Hz - 1kHz
10kΩ
0.01 x |10 x 1.0144 x 10
3
-10x 10
3
|
Z accuracy = (0.08 +
_____________________________
) 0.08%
10 x 10
3
0.005 x |10 x 1.0144 x 10
3
-10x 10
3
|
θ accuracy =
(0.05 +
______________________________
) 0.05
10 x 10
3
Example calculation
1. Impedance (Z=500 Ω) basic accuracy
Measurement conditions: test frequency = 10 kHz, signal level = 1 V,
speed = SLOW2, range = 1 kΩ
From Basic Accuracy Coefficient
Table (0.501V to 1.000V) on the
following pages), basic Z accuracy
coefficients A=0.08, B=0.01.
Inserting these in the calculation expression yields:
Similarly for θ basic accuracy coefficients A = 0.05, B=0.005, and thus:
2. Capacitance(Cs=160 nF) basic accuracy
Measurement conditions: test frequency = 1 kHz, signal level = 1 V,
speed = SLOW2
Test Z and θ.
Measurement range: AUTO
When Z = 1.0144 kΩ, θ = -78.69 ,
test range is 10 kΩ
From Basic Accuracy Coefficient
Table (0.501V to 1.000V) on the
following pages), basic Z accuracy
coefficients A=0.08, B=0.01.
Inserting these in the calculation expression yields:
Similarly for θ basic accuracy coefficients A = 0.05, B=0.005, and thus:

Table of Contents