57
_____________________________________________________________________________________________
4.7 Open Circuit Compensation
______________________________________________________________________________________________
NOTE
NOTE
SPOT compensation for the same frequency as previously
Just press the key. SPOT compensation will be started using the same
frequency which was used for the last episode of SPOT compensation.
SPOT compensation for some new frequency
Input the frequency for compensation using the numeric keypad.
Frequency range which can be set : 42.0 Hz to 5.000 MHz
If you attempt to set a test frequency greater than 5 MHz, it will
automatically be reduced to 5 MHz.
If you attempt to set a test frequency lower than 42 Hz, it will automatically
be set to 42 Hz.
If you make a mistake during input of the numerical value, press the
key to cancel the input so far, and start again.
When you have completed inputting the numerical value for the frequency,
press the appropriate units key
, ,or
so as to confirm the
displayed frequency for compensation.
Until the numerical value for the frequency has been input, the units keys
cannot be used.
3. After the frequency for SPOT compensation has been input as above, press
the key.
Check again - are the test leads definitely out of mutual contact?