116
Input and Measurement Specications
5. Voltage frequency
Range
Display range
(Accuracy guarantee
range)
Accuracy*
2
Minimum sensitivity
voltage
6.000 V
range
60.00 V
range
600.0 V
range
1000 V
range
99.99 Hz
5.00 Hz to 99.99 Hz
(5.00 Hz to 99.99 Hz
*
1
)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
999.9 Hz
40.0 Hz to 999.9 Hz
(40.0 Hz to 999.9 Hz)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
9.999 kHz
0.100 kHz to 9.999 kHz
(0.100 kHz to 9.999 kHz)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
99.99 kHz
1.00 kHz to 50.00 kHz
(1.00 kHz to 50.00 kHz)
±0.1% rdg
±1 dgt
1.800 V 12.00 V 120.0 V 230 V
Over 50.00 kHz to 99.99 kHz
(Over 50.00 kHz to 99.99 kHz)
±0.1% rdg
±1 dgt
3.000 V 24.00 V 240.0 V 400 V
Input impedance, overload protection, coupling type: The same as the AC
voltage RMS value (p. 114 )
Auto range movement threshold: More than 9999 counts for upper range
Less than 900 counts for lower range
If the voltage frequency is displayed in the main display, the AC voltage range
is xed to the 6.000 V range.
If the voltage frequency is displayed in the sub display, the voltage frequency
range is xed to the auto range mode.
*1: The measurement range of 5.00 Hz and above is only for the 6.000 V
range.
The measurement range for other voltage ranges is 40.00 Hz to
99.99 Hz.
*2: ±2 dgt should be added to 20% or less of the range.
HIOKI DT4261A961-00