116
Input and Measurement Specications
5. Voltage frequency
Range
Display range
(Accuracy guarantee
range)
Accuracy*
2
Minimum sensitivity
voltage
6.000 V
range
60.00 V
range
600.0 V
range
1000 V
range
99.99 Hz
5.00 Hz to 99.99 Hz
(5.00 Hz to 99.99 Hz
*
1
)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
999.9 Hz
40.0 Hz to 999.9 Hz
(40.0 Hz to 999.9 Hz)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
9.999 kHz
0.100 kHz to 9.999 kHz
(0.100 kHz to 9.999 kHz)
±0.1% rdg
±1 dgt
0.600 V 6.00 V 60.0 V 100 V
99.99 kHz
1.00 kHz to 50.00 kHz
(1.00 kHz to 50.00 kHz)
±0.1% rdg
±1 dgt
1.800 V 12.00 V 120.0 V 230 V
Over 50.00 kHz to 99.99 kHz
(Over 50.00 kHz to 99.99 kHz)
±0.1% rdg
±1 dgt
3.000 V 24.00 V 240.0 V 400 V
Input impedance, overload protection, coupling type: The same as the AC
voltage RMS value (p. 114 )
Auto range movement threshold: More than 9999 counts for upper range
Less than 900 counts for lower range
If the voltage frequency is displayed in the main display, the AC voltage range
is xed to the 6.000 V range.
If the voltage frequency is displayed in the sub display, the voltage frequency
range is xed to the auto range mode.
*1: The measurement range of 5.00 Hz and above is only for the 6.000 V
range.
The measurement range for other voltage ranges is 40.00 Hz to
99.99 Hz.
*2: ±2 dgt should be added to 20% or less of the range.