41
Installing Wireless Adapter
2.4 Installing Wireless Adapter
When the Z3210 Wireless Adapter (option) is installed to the
instrument, the wireless communications function can be used.
(p. 92)
WARNING
Before removing the battery cover,
disconnect the instrument from the object
under measurement and set the rotary
switch to OFF.
Failure to do so could cause electric shock. When
the instrument is connected to the object under
measurement, the battery contacts are regarded as
high-voltage parts.
After installing or removing the Z3210
Wireless Adapter, attach and then lock the
battery cover.
Using the instrument with the cover removed could
result in bodily injury.
Besides, the cover cannot be secured unless it is
locked.
NOTICE
Before handling the Z3210, touch metal (a door
knob, etc.) to discharge static electricity.
Static electricity may damage the Z3210.