88
DC Voltage Positive/Negative Judgment Function
4.8 DC Voltage Positive/Negative
Judgment Function
When the measured DC voltage value is less than or equal to the
reference value, a buzzer sounds and the backlight turns on in red.
This function can be used to check for any connection errors of the
DC power line.
(Default setting: Disabled)
Reference value: −10 V or less
Measurement function: DC V, AUTO V, or LoZ V
Enabling/disabling the DC voltage positive/negative
judgment function
With the power o, turn the rotary switch while holding down the
MAX/MIN PEAK key.
(Any position)