119
Input and Measurement Specications
10. Electrostatic capacity
Range
Accuracy Measurement current Open circuit voltage
1.000 μF ±1.9% rdg ±5 dgt 10 nA/100 nA/1 μA ±20% 2.0 V DC or less
10.00 μF ±1.9% rdg ±5 dgt 100 nA/1 μA/10 μA ±20% 2.0 V DC or less
100.0 μF ±1.9% rdg ±5 dgt 1 μA/10 μA/100 μA ±20% 2.0 V DC or less
1.000 mF ±1.9% rdg ±5 dgt 10 μA/100 μA/200 μA ±20% 2.0 V DC or less
10.00 mF ±5.0% rdg ±20 dgt 100 μA/200 μA ±20% 2.0 V DC or less
Overload protection: 1000 V DC/1000 V AC or 2 × 10
7
V
Hz, whichever is
lower
(Applied for 1 minute)
Current under short circuit: 300 μA or less
Current under overload
Steady state: 15 mA or less
Transient state: 1.6 A or less
Maximum count for each range: 1100 (1000 for 10.00 mF range)
Auto range movement threshold value: More than 1100 counts for upper
range
Less than 100 counts for lower range