121
Input and Measurement Specications
Peak value (PEAK MAX/PEAK MIN)
Range
Display range
(Accuracy
guarantee range)
Accuracy (only the instrument)
40 Hz ≤ f ≤ 500 Hz 500 Hz < f < 1 kHz
10.00 A
−30.0 A to 30.0 A
(±
2.0
A to ±30.0 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
20.00 A
−60.0 A to 60.0 A
(±
4.0
A to ±60.0 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
50.0 A
−150 A to 150 A
(±
10
A to ±150 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
100.0 A
−300 A to 300 A
(±
20
A to ±300 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
200.0 A
−600 A to 600 A
(±
40
A to ±600 A)
±1.5% rdg ±7 dgt ±2.0% rdg ±7 dgt
500 A*
1
−1500 A to 1500 A
(±
100
A to ±1500 A)
±1.5% rdg ±70 dgt ±2.0% rdg ±70 dgt
1000 A*
1
−1500 A to 1500 A
(±
200
A to ±1500 A)
±1.5% rdg ±70 dgt ±2.0% rdg ±70 dgt
Input impedance, conversion rate, coupling type:
The same as the AC current (clamp sensor) RMS value (p. 120 )
The accuracy does not include the error of the Clamp on Probe.
The maximum input is based on the Clamp on Probe specications.
*1: Minimum resolution 10 A