12
Safety Information
Measurement category III (CAT III)
Applicable to test and measuring circuits connected to the
distribution part of the building’s low-voltage mains installation.
EXAMPLE: Measurements on distribution boards (including
secondary meters), photovoltaic panels, circuit breakers,
wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in the xed installation, and
equipment for industrial use and some other equipment
such as stationary motors with permanent connection to
the xed installation.
Measurement category IV (CAT IV)
Applicable to test and measuring circuits connected at the source of
the building’s low-voltage mains installation.
EXAMPLE: Measurements on devices installed before the main
fuse or circuit breaker in the building installation.
T
Outlet
CAT II
Internal Wiring
Distribution
Panel
Service Entrance
Service Drop
CAT IV
Power Meter
CAT III
Fixed Installation
See: “2.3 Use of Test Leads” (p. 35)
HIOKI DT4261A961-00