7-20
7.2.18 Ambient Temperature Sensor Fault
7.2.19 Viscometer Temperature Sensor Fault
A
mbient Temperature
Sensor Fault
Disconnect the EZJ94 board
CN7 connector and
reconnect it.
Replace the print head.
Replace the EZJ93 board.
Resume the
operation.
Does the fault recur?
Y
es
No
Does the fault recur?
Y
es
No
V
iscometer Temperature
Sensor Fault
Pull out and insert the
connector CN6 of the EZJ93
board.
Replace the thermistor.
Replace the EZJ93 board.
Resume the
operation
.
Does the fault recur?
Y
es
No
Does the fault recur?
Y
es
No