7-39
7.4.9 Viscosity Readings Out of Range
Viscosity Reading
Out of Range
Resume the
operation.
Is it normally measured?
Y
es
No
Perform the viscometer
calibration
Perform the viscometer
operation test on the
Circulation system control
screen.
Replace the proximity sensor.
Disassemble and clean the
viscometer.
Replace the viscometer
Is it normally measured?
Is it normally measured?
Yes
Yes
No
No
Replace the ink.
Is it normally measured?
Yes
No