EasyManua.ls Logo

Hofmann MC10 HS - Changing Settings; Setting Measuring Run Parameters

Hofmann MC10 HS
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MC10 HS Operation - Changing Settings
5 - 13
5.3 Changing Settings
5.3.1 Setting Measuring Run Parameters
ATTENTION
!
All parameters have a direct impact on the measure-
ment data acquisition and, thus, on the measuring
capability of the measurement electronics.
Changes should only be carried out by Hofmann ser-
vice personnel.
5.3.1.1 Measure Settings
Main Menu \ Data Acquisition Setup.
Enter values or change settings:
Data acquisition
Test Mode (No Stop)
- Delay time
Continuous measurement
- Delayed measuring start
Fixed Time then Stop:
- Delay Time
- Measuring time
- Time out
Measurement over a fixed
measuring time set by user.
- Delayed measuring start
- Duration of measurement
- Cancellation of measurement
after specified time
Stability Method:
- Delay Time
-Time out
-Stability limit
-Num of stable readings in
a row
Measuring for as long as it takes
for x successive measured val-
ues to be the stability limit.
The measuring time is limited to
the "Time out" value.
low speed measure profile
or
high speed measure pro-
file
Select for both measurement
profiles:
Master (balancing with averag-
ing)
Production
Bode (measuring with reduced
averaging)

Table of Contents