Insight FD Mini C-Arm Imaging System User Guide
Safety Hazards
11 MAN-09971 Revision 001
Cleaning Safety Almost any liquid spilled into
electronic equipment
automatically becomes a
conductor of electricity.
This system is not waterproof.
If liquid is spilled on the
system, turn the system off
immediately. Do not turn the
system on until the liquid has
dried completely. Do not
splash or spray cleaning
solution on the system.
Clean exterior surfaces of the
Fluoroscan Insight Console,
the flex-arm, and the C-arm
with a soft cloth or disposable
lint-free tissue moistened with
water, isopropanol or
methanol. Wipe surfaces; do
not scrub.
Device Safety
Injury could result if
unapproved devices
are connected to the
system.
Certain devices were tested for
use with the Fluoroscan
System, use only these or
compatible devices.
Do not connect any devices to
the system other than those
described in this manual.
Internally connected devices
should be installed only by
persons trained in the
servicing of this system.
Electrosurgical
Device
This system is shielded from
electrical discharges. However,
there is still a possibility that
an electrosurgical device
discharged directly into the
system could adversely affect
system operation. In addition,
shielding might not always
protect the system from the
amount of energy produced by
a spark-gap type device.
If a strong electrosurgical
surge causes the system to
malfunctions:
Turn the system off
Wait five seconds.
Turn the system on.
If the problem persists after
resetting the system twice,
request service.
Table 2 Safety Hazards
Safety Hazard Cause Course of Action