Selenia User Guide
Appendix ASystem Specifications
MAN-03270 Revision 003 115
A.4 Electrical Input
A.4.1 Tubestand
Mains Voltage
200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation) ±10%, Single Phase,
50/60 Hz ± 5%, permanently wired
Mains Impedance
Maximum line impedance not to exceed 0.20 ohms for
220/230/240 VAC, 0.16 ohms for 200/208 VAC
Maximum Power Consumption
6.5 kVA for 5 second duration
Standby Current
4.0 A (maximum)
Maximum Line Current
35 A for 5 seconds (momentary rating per NEC 660.2)
Recommended Branch Circuit Breaker
Rating
25 A minimum, refer to NEC 660.6
A.4.2 Acquisition Workstation
Input Line Voltage
100/120/200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation), ± 10%
Input Current
8.0 A maximum @ 100/120 VAC
3.5 A maximum @ 200/208/220/230/240 VAC
Frequency
50/60 Hz ±5%
Number of Phases
Single
A.5 Acquisition Workstation Technical Information
Computer Memory
2 GB RAM minimum
Disk capacity
> 60 GB Image Storage
Storage Media
CD-RW Disks
Display Adapter Card
1600 x 1200 matrix minimum 8 bit gray scale display
Display
1600 x 1200 matrix minimum 450 cd/m2 nominal brightness
Network Interface
10/100 Base-T Ethernet
Remote Diagnostics
Internet
Graphical User Interface
X-ray exposure control
Configurable mammographic Workflow
Patient demographics
Brightness and contrast control
Magnification screen
Pixel value readout
QC test tools
System Status Monitoring
Error reporting
Unattended archiving and printing
Radiation Shield Pb equivalence
Rated for a 0.5 mm Pb (lead) equivalence