Selenia User Guide
Appendix A—System Specifications
MAN-03270 Revision 003 115
A.4 Electrical Input
A.4.1 Tubestand
200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation) ±10%, Single Phase,
50/60 Hz ± 5%, permanently wired
Maximum line impedance not to exceed 0.20 ohms for
220/230/240 VAC, 0.16 ohms for 200/208 VAC
Maximum Power Consumption
6.5 kVA for 5 second duration
35 A for 5 seconds (momentary rating per NEC 660.2)
Recommended Branch Circuit Breaker
Rating
25 A minimum, refer to NEC 660.6
A.4.2 Acquisition Workstation
100/120/200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation), ± 10%
8.0 A maximum @ 100/120 VAC
3.5 A maximum @ 200/208/220/230/240 VAC
A.5 Acquisition Workstation Technical Information
1600 x 1200 matrix minimum 8 bit gray scale display
1600 x 1200 matrix minimum 450 cd/m2 nominal brightness
Configurable mammographic Workflow
Patient demographics
Brightness and contrast control
Magnification screen
Pixel value readout
QC test tools
System Status Monitoring
Error reporting
Unattended archiving and printing
Radiation Shield Pb equivalence
Rated for a 0.5 mm Pb (lead) equivalence