Selenia User Guide
Appendix A—System Specifications
114 MAN-03270 Revision 003
A.1.3 Selenia Value Console
Figure 83: Value Console Dimensions
A
Height
191.3 cm (75.3 inches)
B
Width
101.2 cm (39.8 inches)
C
Depth
60.5 cm (23.8 inches)
Weight
163.3 kg (360 lb) (maximum)
A.2 Operating Environment
A.2.1 General Operating Conditions
Temperature Range
20 °C (68° F) to 30 °C (86° F)
Maximum rate of temperature change
<10 ºC/hr
Relative Humidity Range
10% to 80% non-condensing
BTU output
Typical Range 1700–2500 BTU/hr
A.3 Storage Environment
A.3.1 Tubestand
Temperature Range
-25 °C (-13° F) to +60 °C (140° F)
Humidity
Zero to 95% humidity—non-condensing (not packaged for
outdoor storage)
A.3.2 Image Receptor
Storage Temperature Range
10 ºC (50° F) to 40 ºC (104° F)
Maximum rate of temperature change
<15 ºC/hr
Storage Humidity Range
10% to 80% humidity— non-condensing (not packaged
for outdoor storage)

Table of Contents