Selenia User Guide
Appendix BThe Mobile Selenia
MAN-03270 Revision 003 129
B.2 Safety Conditions and Other Precautions
Warning:
The radiation shield is not approved for mobile use
and is not provided. The coach manufacturer must
provide adequate shielding.
Caution:
When shore power is unavailable, mobile power sources
that provide equivalent performance may be employed.
(See Mobile Specifications.) Proper system function and
performance can only be ensured if continuous true
sinusoidal VAC power is supplied per the system power
input specifications and loading characteristics.
Intermittently, the power source must provide 65 Amps at
208 VAC for minimum of 5 seconds, and 4 Amps
maximum continuous otherwise. This load must be
supported once every 30 seconds. In the event of shore or
mobile power service interruption, the UPS must be
capable of providing the operational power described
above for a minimum of 4 minutes. Acquisition
Workstation and Gantry power must be fed on separate
dedicated circuits. The use of an uninterruptible power
supply with active line conditioner is recommended on
each power circuit. Accordingly, all ancillary mobile coach
power should be distributed by other circuits. The
electrical installation must be verified to meet system
power input specifications and IEC 60601-1 safety
requirements after initial installation and upon each
relocation of the mobile coach.
Caution:
The temperature and humidity inside the vehicle must be
maintained at all times. Do not allow environmental
conditions to exceed stated specifications when the unit is
not in use.
Caution:
For permanent archive, the Selenia must be attached to a
PACS system or connected to a suitable hard-copy printer
when in use.

Table of Contents