Selenia User Guide
Appendix B—The Mobile Selenia
130 MAN-03270 Revision 003
B.3 Mobile Specifications
B.3.1 Shock and Vibration Limits
Not greater than 0.35 G (2 Hz to 200 Hz), as measured at
mounting point of system to coach.
Not greater than 1.0 G (½ sine pulse), as measured at mounting
point of system to coach. An "air ride" coach suspension is
recommended.
B.3.2 Coach Environment
B.3.2.1 Operating Environment
20 °C (68° F) to 30 °C (86° F)
Maximum Rate of Temperature Change
10% to 80% non-condensing
Typical Range 1700–2500 BTU/hr
B.3.2.2 Non-operating/Transit Environment
15 °C (59° F) through 35 °C (95° F)
Maximum Rate of Temperature Change
10% to 80% non-condensing
B.3.3 Electrical Input
B.3.3.1 Tubestand
200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation) ±10%, Single Phase,
50/60 Hz ± 5%, permanently wired
Maximum line impedance not to exceed 0.20 ohms for
220/230/240 VAC, 0.16 ohms for 200/208 VAC
Maximum Power Consumption
6.5 kVA for 5 second duration
35 A for 5 seconds (momentary rating per NEC 660.2)
Recommended Branch Circuit Breaker
Rating
25 A minimum, refer to NEC 660.6
B.3.3.2 Acquisition Workstation
100/120/200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation), ± 10%
8.0 A maximum @ 100/120 VAC
3.5 A maximum @ 200/208/220/230/240 VAC