Selenia User Guide
Appendix BThe Mobile Selenia
130 MAN-03270 Revision 003
B.3 Mobile Specifications
B.3.1 Shock and Vibration Limits
Vibration Limit
Not greater than 0.35 G (2 Hz to 200 Hz), as measured at
mounting point of system to coach.
Shock Limit
Not greater than 1.0 G (½ sine pulse), as measured at mounting
point of system to coach. An "air ride" coach suspension is
recommended.
B.3.2 Coach Environment
B.3.2.1 Operating Environment
Temperature Range
20 °C (68° F) to 30 °C (86° F)
Maximum Rate of Temperature Change
<10 ºC/hr
Relative Humidity Range
10% to 80% non-condensing
BTU output
Typical Range 17002500 BTU/hr
B.3.2.2 Non-operating/Transit Environment
Temperature Range
15 °C (59° F) through 35 °C (95° F)
Maximum Rate of Temperature Change
<15 °C/hr
Relative Humidity Range
10% to 80% non-condensing
B.3.3 Electrical Input
B.3.3.1 Tubestand
Mains Voltage
200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation) ±10%, Single Phase,
50/60 Hz ± 5%, permanently wired
Mains Impedance
Maximum line impedance not to exceed 0.20 ohms for
220/230/240 VAC, 0.16 ohms for 200/208 VAC
Maximum Power Consumption
6.5 kVA for 5 second duration
Standby Current
4.0 A (maximum)
Maximum Line Current
35 A for 5 seconds (momentary rating per NEC 660.2)
Recommended Branch Circuit Breaker
Rating
25 A minimum, refer to NEC 660.6
B.3.3.2 Acquisition Workstation
Input Line Voltage
100/120/200/208/220/230/240 VAC nominal, true sinusoidal,
(tap selectable at installation), ± 10%
Input Current
8.0 A maximum @ 100/120 VAC
3.5 A maximum @ 200/208/220/230/240 VAC
Frequency
50/60 Hz ±5%
Number of Phases
Single