Selenia User Guide
Chapter 4Startup, Functional Tests, and Shutdown
34 MAN-03270 Revision 003
Table 5:
Monthly Control Function Tests
Function
Control
Test
Clockwise C-arm Rotation
C-arm and Gantry Rotation
Switches
Press the button and confirm the
following actions:
C-arm rotates
Angle display changes
Rotation stops when you
release the button.
Press and hold a Center
button and confirm C-arm
moves to zero degree
position.
Compression Down
1. Press the button.
2. Confirm the compression brake
engages and the light field lamp
illuminates.
3. Confirm the Compression Down
movement stops:
When you release the
button.
At the compression down
force limit.
At the lower compression
travel limit.

Table of Contents