Selenia User Guide
Chapter 4Startup, Functional Tests, and Shutdown
MAN-03270 Revision 003 35
Table 5:
Monthly Control Function Tests
Function
Control
Test
Compression Up
1. Press the button.
2. Confirm the compression brake
does not release.
3. Confirm the Compression Up
movement stops:
When the Compression
device reaches the upper
compression travel limit.
On release of the button.
Compression Release
1. Press the button.
2. Confirm the following actions:
The compression motor
brake releases.
The compression device lifts
approximately 10 cm.
Note The system disables all compression
release functions when a localization
paddle is installed, or system defaults are
configured for manual compression
release.
C-arm Up
1. Press the button.
2. Confirm the C-arm Up
movement stops:
When you release the
button.
When the C-arm reaches the
upper travel limit.
3. Apply more than 58 N (13 lb)
compression and confirm the
C-arm does not move.