Selenia User Guide
Chapter 4Startup, Functional Tests, and Shutdown
36 MAN-03270 Revision 003
Table 5:
Monthly Control Function Tests
Function
Control
Test
C-arm Down
1. Press the button.
2. Confirm the C-arm Down
movement stops:
When you release the
button.
When the C-arm reaches the
lower travel limit.
Light Field Lamp
1. Press the button. Confirm the
light turns on for a short time.
2. Confirm the light field lamp
turns on when a Compression
Down button is activated.
Collimator Override
1. Press the light field button then
the collimator button.
2. Confirm the Collimator moves
to the next field size.
3. Repeat the steps 1 and 2 until
you move through all Collimator
field sizes.