Monthly Control Function Tests
1. Press the button.
2. Confirm the C-arm Down
movement stops:
• When you release the
button.
• When the C-arm reaches the
lower travel limit.
1. Press the button. Confirm the
light turns on for a short time.
2. Confirm the light field lamp
turns on when a Compression
Down button is activated.
1. Press the light field button then
the collimator button.
2. Confirm the Collimator moves
to the next field size.
3. Repeat the steps 1 and 2 until
you move through all Collimator
field sizes.