EasyManuals Logo

Honeywell IntelliDoX User Manual

Honeywell IntelliDoX
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
INTELLIDOX DOCKING MODULE USER MANUAL || TROUBLESHOOTING
HONEYWELL PAGE 110 OF 119
5. Install the new cradle. See Replace Detector Cradle and Calibration
Insert.
6. Connect the power cord to the AC outlet.
The module firmware is not updated when a new firmware update file is
transferred from USB flash drive.
The file name is incorrect.
The firmware update file name is not the same as the downloaded file name.
Rename the file name to match the name of the downloaded firmware update
file, or download a new firmware update file. See Update Firmware.
The file is not in the System folder on the USB flash drive.
Create a folder called System on the USB flash drive. Move the firmware update
file to the System folder.
The firmware on the module is the same version as the firmware update.
Verify that the firmware file on the USB flash drive is the most recent firmware
version. See Update Firmware.
The module firmware is not updated on modules that are connected in a
group.
Modules in the group are busy.
Remove detectors from all modules in the group. The firmware update is not
installed when modules in the group are busy (for example, when a detector is in
a module).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell IntelliDoX and is the answer not in the manual?

Honeywell IntelliDoX Specifications

General IconGeneral
Data LoggingYes
CalibrationAutomatic
Bump TestingAutomatic
Instrument ManagementYes
Input Voltage100-240 VAC, 50/60 Hz
Enclosure RatingIP20
ConnectivityEthernet, USB
Operating Temperature0°C to 40°C (32°F to 104°F)
Humidity Range10% to 90% (non-condensing)
Communication ProtocolModbus TCP/IP

Related product manuals