INTELLIDOX DOCKING MODULE USER MANUAL || IMPORTANT SAFETY
INFORMATION: READ FIRST
HONEYWELL PAGE 4 OF 119
Important Safety Information: Read First
To ensure personal safety, read all Safety
Information and Warnings before using this
equipment.
Use this equipment only as specified by the
manufacturer. Failure to do so may impair
protection provided by the equipment.
The safety and security of any system or network
incorporating this equipment and its accessory
components is the responsibility of the assembler
of the system.
Follow all required National Electric Codes (NEC)
and safety standards.
This IntelliDoX module is intended for use as a
stand-alone docking module, or as a component
in a connected group of up to five IntelliDoX
modules.
If the equipment is damaged or parts are missing,
contact Honeywell immediately.
Substitution of components may impair the safety
of the equipment.
Do not attempt to disassemble, adjust, or service
the equipment unless instructions are provided to
perform a procedure, or a component is listed as
a user-replaceable component in the user
manual. Use only Honeywell replacement
components.
To eliminate the risk of electrical shock,
disconnect the equipment when cleaning or
performing maintenance.
This equipment uses potentially harmful gas for
calibrations. The equipment must be attached to
a venting system or be used in a well-ventilated
area.
To prevent gas leaks, the end plate must be
attached and locked with the latch arm lock
before connecting power supply and/or
connecting gas cylinders. The end plate must
remain securely latched at all times during
operation.
Perform calibrations and bump tests only in a
normal atmosphere (20.9% v/v O
2
) that is free of
hazardous gas.
The maximum recommended calibration tubing
length is 10 meters (33 feet). Use of tubing that is
longer than the recommended length may result
in inaccurate or failed compliance tests or
calibrations.
The maximum recommended exhaust tubing
length is 15 meters (49.5 feet).
Ensure that the inlet filter is clean.
The use of calibration gas cylinders other than
those specified by Honeywell may result in
unsafe calibration or possible non-recoverable
failure of the equipment, and will invalidate the
warranty.
Do not use calibration gas cylinders past their
expiration date.
Ensure that all calibration gas cylinders are in
good condition.
Ensure that all calibration gas cylinders contain
enough gas.
Ensure the exhaust line is not connected to a
negative pressure system.
A demand flow regulator must be used with all
calibration gas cylinder connections. The module
inlet should not be pressurized.
Failed tests can result if inlets are not set up
correctly.
Do not expose the equipment to electrical shock
or severe and/or continuous mechanical shock.
Do not immerse the equipment in liquids.
Products may contain materials that are
regulated for transportation under domestic and
international dangerous goods regulations.
Return product in compliance with appropriate
dangerous goods regulations. Contact freight
carrier for further instructions.
To prevent the corruption or loss of data and/or
software and/or firmware, do not deactivate the
equipment while performing datalog transfers,
bump tests, calibrations or other operations.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell IntelliDoX and is the answer not in the manual?

Honeywell IntelliDoX Specifications

General IconGeneral
Data LoggingYes
CalibrationAutomatic
Bump TestingAutomatic
Instrument ManagementYes
Input Voltage100-240 VAC, 50/60 Hz
Enclosure RatingIP20
ConnectivityEthernet, USB
Operating Temperature0°C to 40°C (32°F to 104°F)
Humidity Range10% to 90% (non-condensing)
Communication ProtocolModbus TCP/IP