Honeywell Sensing and Control     31
SC 2000, SC2001, SC3004
 Rev. G,  008-0608-00
10.5.4 CALIBRATION TYPE Menu Item
This chooses the type of calibration technique to be used. There 
are three methods that can be used to calibrate the transducer to 
the Input Channel. Each has advantages and disadvantages as 
described in the table below
Table 10-: Comparison of Calibration Types
Shunt Calibration
mV/V Calibration
2-, 3-, or 5-point 
Known Load Cali-
bration
RECOMMENDED...
...for most applications no yes no
...when frequently 
swapping transducers
yes yes no
...when best possible 
accuracy required
no no yes
PROs and CONs
Relative accuracy good better best
Requires actual, 
calibrated loads to be 
applied
no no yes
Automatically cali-
brates with Signature 
Calibration transduc-
ers 
yes yes no
Max. distance of trans-
ducer from instrument 
w/out sense wires
15 ft [5 m] 50 ft  
[15 m]
300 ft  
[100 m]
Max. distance of trans-
ducer from instrument 
with sense wires
15 ft [5 m] 100 ft  
[30 m]
300 ft  
[100 m]
CALIBRATION TECHNIQUE
Zero-scale point from... applied 
load
applied 
load
applied 
load
Full-scale point from...  transducer 
output 
when shunt 
resistor 
applied
internal 
reference
applied 
load
Linearity correction 
from...
data 
stored in 
Signature 
Module (if 
available)
data 
stored in 
Signature 
Module (if 
available)
2-point: 
none;
3- or 
5-point: 
applied 
load
It is important to know your application in order to select the 
appropriate calibration type. mV/V (“millivolt-per-volt”) Calibra-
tion is recommended for most applications because it doesn’t 
require known loads and it allows easy swapping of transducers 
equipped with Signature Calibration. When the best possible ac-
curacy is required, Known-Load Calibration is recommended.
The choices for this menu item are:
• “TYPE= SHUNT CAL” means Shunt Calibration. First, you are 
prompted to apply the load entered in the “ZERO SCALE VAL-
UE” register. Next, the instrument inserts a known, precision 
resistor into the transducer circuit, which causes a predict-
able apparent signal. The instrument then takes a reading 
and adjusts itself using the “SHUNT CAL VALUE” register.
• “TYPE= MV/V CAL” means Millivolt-per-Volt Calibration. First, 
you are prompted to apply the load entered in the “ZERO 
SCALE VALUE” register. Next, the instrument takes a reading 
from an accurate, internal mV/V reference. The instrument 
then uses the “FULL SCALE MV/V” and “FULL SCALE VALUE” val-
ues in mathematical calculations to scale its tracking values.
• “TYPE= 2 POINT CAL” means 2-Point Known Load Calibra-
tion. You are prompted to apply the loads to the transducer 
that were entered in the “KNOWN POINT 1/2” and “KNOWN POINT 
2/2” registers. This technique assumes that the transducer is 
linear, so the usual loads used are zero scale and full scale.
• “TYPE= 3 POINT CAL” means 3-Point Known Load Calibration. 
You are prompted to apply the loads to the transducer that 
were entered in the “KNOWN POINT 1/3” and “KNOWN POINT 2/3” 
and “KNOWN POINT 3/3” registers. This technique can be used 
to compensate for the non-linearity in the transducer. The 
usual loads used are zero scale, half scale and full scale, but 
you are not restricted to these loads.
• “TYPE= 5 POINT CAL” means 5-Point Known Load Calibration. 
You are prompted to apply the loads to the transducer that 
were entered in the “KNOWN POINT 1/5”, “KNOWN POINT 2/5”, 
“KNOWN POINT 3/5”, “KNOWN POINT 4/5” and “KNOWN POINT 
5/5” registers. This technique can be used to compensate 
for the non-linearity in the transducer. The usual loads used 
are zero scale, one-quarter scale, half scale, three-quarters 
scale and full scale, but you are not restricted to these loads.
The selection of the calibration type affects what menu items ap-
pear in the CALIBRATION DATA sub-menu.