Honeywell Sensing and Control     69
SC 2000, SC2001, SC3004
 Rev. G,  008-0608-00
14.7 Troubleshooting
14.7.1 Error Messages
See “Error Messages” on page 73 for information relating to error 
messages.
14.7.2 Common Problems and Solutions
 
Analog Output Incorrect
Make certain of the type of Analog Output (voltage or current) 
that the channel is equipped with; see “Identifying the Output 
Type” on page 68.
Use the DIAGNOSTICS -> DAC FULL SCALE and DIAGNOSTICS -> DAC 
ZERO SCALE menu items to force the Analog Output to a known 
output. Then, adjust your readout device, panel meter, PLC or 
data acquisition system to match.
Check the output selection jumpers; see “Output Selection” on 
page 68.
Check the “DAC SETUP -> DAC. ZERO-SCALE” and “DAC SETUP -> 
DAC FULL-SCALE” menu items.
Sensitivity to EMI/RFI
To obtain maximum immunity to electromagnetic or radio fre-
quency interference, make certain that the shield of the cable is 
connected to ground/earth only at the readout device. The shield 
of the cable should not be connected to the instrument.