34    Honeywell Sensing and Control
Signal Conditioning, Self Calibrating Digital Indicators
  Rev. G, 008-0608-00
If the CALIBRATION TYPE is Shunt Calibration...
• ThedisplaywillreadDOING SHUNT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as ZERO SCALE VALUE and DISPLAY. UNITS. When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillthenreadWORKING and calibration will take 
place. Upon completion, the display will indicate DONE and 
the instrument will return to the RUN mode.
If the CALIBRATION TYPE is mV/V Calibration...
• ThedisplaywillreadDOING MV/V CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as ZERO SCALE VALUE and DISPLAY. UNITS). When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillthenreadWORKING and calibration will take 
place. Upon completion, the display will indicate DONE and 
the instrument will return to the RUN mode.
If the CALIBRATION TYPE is 2-Point Known Load Calibration...
• ThedisplaywillreadDOING 2POINT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as KNOWN POINT 1/2 and DISPLAY. UNITS). When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillreadWORKING, then APPLY 01000 UNIT (where 
01000 and UNIT are previously entered as KNOWN POINT 2/2 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER].
• Uponcompletion,thedisplaywillindicateDONE and the 
instrument will return to the RUN mode.
If the CALIBRATION TYPE is 3-Point Known Load Calibration...
• ThedisplaywillreadDOING 3POINT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as KNOWN POINT 1/3 and DISPLAY. UNITS). When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillreadWORKING, then APPLY 00500 UNIT (where 
00500 and UNIT are previously entered as KNOWN POINT 2/3 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER]. 
• ThedisplaywillreadWORKING, then APPLY 01000 UNIT (where 
01000 and UNIT are previously entered as KNOWN POINT 3/3 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER]. 
• Uponcompletion,thedisplaywillindicateDONE and the 
instrument will return to the RUN mode.
If the CALIBRATION TYPE is 5-Point Known Load Calibration...
• ThedisplaywillreadDOING 5POINT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as KNOWN POINT 1/5 and DISPLAY. UNITS). When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillreadWORKING, then APPLY 00250 UNIT (where 
00250 and UNIT are previously entered as KNOWN POINT 2/5 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER].
• ThedisplaywillreadWORKING, then APPLY 00500 UNIT (where 
00500 and UNIT are previously entered as KNOWN POINT 3/5 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER].
• ThedisplaywillreadWORKING, then APPLY 00750 UNIT (where 
00750 and UNIT are previously entered as KNOWN POINT 4/5 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER].
• ThedisplaywillreadWORKING, then APPLY 01000 UNIT (where 
01000 and UNIT are previously entered as KNOWN POINT 5/5 
and DISPLAY. UNITS). When you have applied this load to 
the transducer, press [ENTER].
• ThedisplaywillnowreadRETURN PNTS.=NO. If you press 
[ENTER], the display will indicate DONE and the instrument 
will return to the RUN mode. If you use the [UP] button and 
[ENTER] to select RETURN POINTS=YES, you may now apply 
the same loads again but in reverse order. This will allow the 
instrument to measure and partially eliminate any hysteresis 
that may be present in the transducer. After the loads have 
been applied again, the display will indicate DONE and the 
instrument will return to the RUN mode.
10.5.7 DAC SETUP Sub-Menu
This sub-menu contains four items that control the Digital-to-Ana-
log Converter (DAC) output of the channel.
DAC. CHANNEL Menu Item
This chooses which channel will drive the DAC output. Normally, 
the DAC located on a particular channel will be driven by that 
channel, but that need not be the case. For example, if several 
channels were assigned to load cells on a weighing platform, the 
sum of all channels could (mathematically) drive one channel’s 
DAC output.
If the DAC is set-up to be driven by its own channel’s tracking 
value, the Analog Output will have the frequency response speci-
fied by the channel’s FREQ. RESPONSE menu item. Otherwise, 
the Analog Output will operate more slowly.