28    Honeywell Sensing and Control
Signal Conditioning, Self Calibrating Digital Indicators
  Rev. G, 008-0608-00
10.3 Calibration Procedure
If you are not familiar with operating the instrument in the SETUP 
menu mode, see “SETUP Menu mode” on page 7. A listing of all 
menu items is given in “Setup Menu Reference” on page 77.
• Step1:Wirethetransducertothechannel’sconnector.
  See “Wiring” on page 27 for details.
• Step2:EntertheCALIBRATION TYPE. 
  There are three methods that can be used to calibrate the 
transducer to the Input Channel. Each has advantages and 
disadvantages as described in “CALIBRATION TYPE Menu 
Item” on page 31. It is important to know your application in 
order to select the appropriate calibration type.
• Step3:EntertheCALIBRATION DATA.
  - If your transducer has Signature Calibration, you don’t 
need to enter these values as they are entered automatically. 
  - If your transducer was ordered and shipped along with 
the instrument, you don’t need to enter these values as they 
have been entered at the factory.
  - Otherwise, consult the Certificate of Calibration for the 
transducer when entering information in the CALIBRATION 
DATA sub-menu.
• Step4:Performthecalibration.
  Otherwise, use the CALIBRATE menu item to start the calibra-
tion process. You will be prompted to apply loads to the 
transducer as required.
10.4 Specications
TRANSDUCER INPUT
Transducer type  full-bridge, strain gage 
Excitation voltage 5 Vdc or 10 Vdc, short circuit pro-
tected with sensing
Transducer full-scale 
output
0.5 to 11.0 mV/V @ 5V excitation
0.5 to 5.5 mV/V @ 10 V excitation
Amplifier gain selection automatic 
Calibration type shunt, mV/V or 2-, 3- or 5-point 
known load
Differential Input Volt-
age
±55mV(max.)
A/D Converter 24-bit Sigma-Delta
Low-pass filter digital, 24-tap FIR
Resolution and Fre-
quency Response
see “FREQ. RESPONSE Menu Item” 
on page 29
INSTRUMENT-ONLY  ACCURACY
(Sense wires used; frequency response setting <16 Hz; Linearity, repeatability, 
and hysteresis)
mV/V calibration ±0.1%FS
Known-load calibration ±0.01%FS
AUXILLARY INPUTS
Quantity 2
Type momentary contact closure
Response time < 5 ms
Field-selectable func-
tions
tare on, tare off, peak/valley clear, 
peak/valley hold, track hold
ANALOG OUTPUTS
Output voltage range 5,±5,10or±10Vdc 
(field selectable)
Output current range 
(optional current output 
channels)
4 mA to 20 mA
Source any channel’s track, peak, or valley 
value
Isolation 500 V
Resolution 13 bits
Frequency response same as input when driven by the 
same channel’s tracking data