42    Honeywell Sensing and Control
Signal Conditioning, Self Calibrating Digital Indicators
  Rev. G, 008-0608-00
DISPLAY. UNITS Menu Item
Specifies the four character label that is displayed to the right of 
the channel’s values.
NOTICE
This menu item doesn’t change the mathematical scaling of 
the channel’s values; that can be changed by altering the 
“CALIBRATION DATA -> KNOWN POINT X/Y” menu items and then 
performing a re-calibration.
When a character position is flashing press the [UP] or [DOWN] 
button to change the character. Press [ENTER] to advance to the 
next character.
DISPLAY. AVERAGE Menu Item
Controls the speed with which the channel’s display values will 
update. Display averaging does not affect the channel’s analog 
output or its peak/valley detection, which will proceed at the 
speed selected by the ”OPERATION -> FREQ. RESPONSE” menu item. 
The choices are:
• “ON” means that the display will update four times each sec-
ond. The channel’s values will be averaged for 1/4 second, 
then displayed.
• “OFF” means that the channel’s display will update as quickly 
as possible.
11.5.3 AUXn FUNCTION Menu Items
The AUX1 FUNCTION and AUX2 FUNCTION menu items determine 
what happens when the Auxiliary Function pins (labeled as 
“AUX1” and “AUX2”) on the channel’s connector are activated. 
These pins are “activated” when they are connected to the 
DGND pin. The choices are:
• “DISABLED” means that activating the pin does nothing.
• “TRACK HOLD” means that the tracking, peak and valley values 
will not be updated.
• “HIGH/LOW HOLD” means that the peak and valley values will 
not be updated.
• “HIGH/LOW CLEAR” means that the peak and valley values are 
reset. 
• “TARE ON” activates the channel’s Tare function (resets it to 
zero).
• “TARE OFF” deactivates the channel’s Tare function.
As the Auxiliary Function pins are not isolated, it is recommended 
that a push-button switch or relay is used to connect these pins 
to the DGND pin.
11.5.4 CALIBRATION TYPE Menu Item
The choices for this menu item are:
• “TYPE= 2 POINT CAL” means 2-Point Known Displacement 
Calibration. You are prompted to apply the displacements 
to the transducer that were entered in the “KNOWN POINT 1/2” 
and “KNOWN POINT 2/2” registers. This technique assumes 
that the transducer is linear throughout the entire stroke, so 
the usual displacements used are zero scale and full scale.
• “TYPE= 3 POINT CAL” means 3-Point Known Displacement 
Calibration. You are prompted to apply the displacements 
to the transducer that were entered in the “KNOWN POINT 1/3” 
and “KNOWN POINT 2/3” and “KNOWN POINT 3/3” registers. This 
technique can be used to compensate for the different sen-
sitivities that an LVDT may have on each side of its electrical 
null. The usual displacements used are zero scale, half scale 
and full scale, but you are not restricted to these displace-
ments.
• “TYPE= 5 POINT CAL” means 5-Point Known Displacement 
Calibration. You are prompted to apply the displacements to 
the transducer that were entered in the “KNOWN POINT 1/5”, 
“KNOWN POINT 2/5”, “KNOWN POINT 3/5”, “KNOWN POINT 4/5” 
and “KNOWN POINT 5/5” registers. This technique can be used 
to compensate for the non-linearity in the transducer as well 
as the different sensitivities on each side of its electrical null. 
The usual displacements used are zero scale, one-quarter 
scale, half scale, three-quarters scale and full scale, but you 
are not restricted to these displacements.
The selection of the calibration type affects what menu items ap-
pear in the CALIBRATION DATA sub-menu. 
11.5.5. CALIBRATION DATA Sub-Menu
This sub-menu enters the values that will be used by the instru-
ment when it calibrates itself to the transducer. The menu items 
that are displayed will change according to what the CALIBRATION 
TYPE is set for.
FULL SCALE VRMS Menu Item
The transducer output when its full-scale displacement is applied 
to it, in Volts RMS (root-mean-squared) assuming 3 VRMS excita-
tion. This menu item is used by the instrument to set gain of its 
amplifier circuit. This value is normally found on the transducer’s 
Certificate of Calibration issued by its manufacturer.